Method Validation for Hexavalent Chromium in Solid Waste

1.Validation Overview

Project Name: Method Validation for Hexavalent Chromium in Solid Waste

Reference Standard:HJ 687—2014《Solid Waste – Hexavalent Chromium, Alkaline Digestion / FAAS》

Instrument: AA-1800C Atomic Absorption Spectrophotometer

Organization:Shanghai Macylab Instrument Inc.

Date: March 05,2026

Validator: He Ke

Reviewer: Jiang Ying

Report number:MACY-2026-0305

2.Instrument Conditions

2.1 Parameters

ParameterValue
Parameter357.9 nm(Cr)
Spectral Bandwidth0.4 nm
Lamp Current6 mA
Burner Height6 mm
Burner50 mm Ti head
Acetylene Flow1.6 L/min
Air Flow8.0 L/min
Integration Time3 s
ReadingPeak Height
Replicates3 times, average

2.2 Performance

IndexReq.ResultConclusion
Static Baseline Drift≤±0.002A/30min±0.0010A/30minPass
Dynamic Baseline Drif≤±0.005A/30min±0.0025A/30minPass
RSD≤0.5%0.28%Pass
Wavelength Accuracy≤0.15nm0.06nmPass
Characteristic Conc≤0.04μg/mL/1%0.028μg/mL/1%Pass

3.Calibration Curve Validation

3.1 Standard Solution Prep

Certified Cr(VI) stock (1000 mg/L), serially diluted by 0.1 mol/L NaOH:

Std NoConc.(mg/L)Method
00.00Blank(0.1 mol/L NaOH)
11.00Serial dilution
22.00Serial dilution
34.00Serial dilution
48.00Serial dilution
516.00Serial dilution
632.00Serial dilution

3.2 Curve Data

Conc.(mg/L)A1A2A3Avg
0.000.00100.00080.00090.0009
1.000.03520.03580.03550.0355
2.000.07080.07120.07100.0709
4.000.14150.14200.14180.1418
8.000.28320.28380.28350.2835
16.000.56580.56650.56620.5662
32.001.12851.12921.12881.1288

3.3 Fitting

ParamResult
Equationy = 0.0352x + 0.0005
R(r)0.9999
Intercept0.0005
Slope0.0352

Criterion: r ≥ 0.995

Conclusion:r = 0.9999,compliant

4. MDL & LOQ Validation

4.1 MDL Test

Per HJ 687-2014: 2.5g sample, final vol 100mL; MDL=2mg/kg, LOQ=8mg/kg.

Test:0.5 mg/L soln, 7 replicates. MDL = t×S, t=3.143(df=6)

No.AbsConc(mg/L)
10.01780.505
20.01720.488
30.01820.518
40.01750.496
50.01800.511
60.01700.482
70.01760.499

Result:

Mean:0.500 mg/L

S:0.013 mg/L

MDL = 3.143 × 0.013 = 0.041 mg/L(soln)

Converted to solid waste:0.041 × 100 / 2.5 = 1.64 mg/kg

4.2 Lower limit determination validation

ItemReqResultConclusion
MDL2 mg/kg1.64 mg/kgSuperior
LOQ8 mg/kg6.56 mg/kgSuperior

Rule: LOQ=4×MDL

Conclusion: MDL & LOQ meet HJ 687-2014.

5. Precision Validation

5.1 Test Scheme

Per HJ 687-2014: low/med/high levels, 6 replicates each.

Conc LevelSpike(mg/kg)Soln Conc(mg/L)
Low8.00.20
Mid40.01.00
High160.04.00

5.2 Precision Results

Low(8.0 mg/kg)

No.Result(mg/kg)
18.12
28.25
38.18
48.22
58.15
68.20

Mean:8.18 mg/kg

S:0.048 mg/kg

RSD = 0.59%

Med(40.0 mg/kg)

No.Result(mg/kg)
140.15
240.28
340.22
440.18
540.25
640.20

Mean:40.21 mg/kg

S:0.052 mg/kg

RSD = 0.13%

High(160.0 mg/kg)

No.Result(mg/kg)
1160.35
2160.28
3160.42
4160.30
5160.38
6160.32

Mean:160.34 mg/kg

S:0.053 mg/kg

RSD = 0.03%

5.3 Precision Conclusion

Conc LevelSpike(mg/kg)RSD(%)ReqResult
Low8.00.59≤10%Pass
Med40.00.13≤5%Pass
High160.00.03≤5%Pass

HJ687-2014: RSD ≤10%(low), ≤5%(med/high).

Conclusion: 3 levels pass, precision qualified

6.Accuracy Validation(Recovery)

6.1 Scheme

Test by spike recovery using solid waste CRM.

Reference Material Information:

Reference Material Name:Standard reference material for the analysis of hexavalent chromium in industrial solid waste

CRM:RMH-G015

Cert:18.5 mg/kg

Uncert:±1.5 mg/kg

6.2 Spike RecoveryLow spike(8.0 mg/kg)

No.Blank(mg/kg)Spiked(mg/kg)Rec(%)
10.158.22101.6
20.128.18101.3
388.28102.0
40.148.20101.4
50.68.25101.8
60.138.19101.3

Avg Rec:101.5%

Rec Range:101.3% ~ 102.0%

Med spike(40.0 mg/kg)

No.Blank(mg/kg)Spiked(mg/kg)Rec(%)
10.1840.3299.8
20.1540.2599.6
30.2040.38100.0
40.1740.3099.9
50.1940.3599.9
60.1640.2899.7

Avg Rec:99.8%

Rec Range:99.6% ~ 100.0%

High spike(160.0 mg/kg)

No.Blank(mg/kg)Spiked(mg/kg)Rec(%)
10.22160.4599.9
20.18160.3899.8
30.25160.52100.0
40.20160.4299.9
50.23160.4899.9
60.19160.4099.8

Acg Rec:99.9%Rec

Range:99.8% ~ 100.0%

6.3 CRM Verification

Test with solid waste CRM RMH-G015 (cert:18.5 mg/kg:

No.Result(mg/kg)
118.42
218.55
318.48
418.52
518.50
618.46

Mean:18.48 mg/kg

Cert:18.5 mg/kg

RE:(18.48 – 18.5) / 18.5 × 100% = -0.11%

RSD:0.32%

Criterion: Measured value within certified uncertainty

Conclusion: 18.48 mg/kg falls in 18.5±1.5 mg/kg, pass.

6.4 Accuracy Conclusion

Conc LevelSpike(mg/kg)Avg Rec(%)ReqResult
Low8.0101.585%~115%Pass
Med40.099.890%~110%Pass
High160.099.990%~110%Pass
CRM18.5(cert)RE-0.11%≤10%Pass

Conclusion: Spike recovery & CRM test meet requirements, method accuracy qualified

7. Actual Sample Validation

7.1 Actual Sample Information

Sample No.TypeLocationDate
SW-2026-001Ind.Solid WasteShanghai Electroplating Plant2026-02-20
SW-2026-002Ind. Solid WasteShanghai Chemical Plant2026-02-22
SW-2026-003Gen. Solid WasteShanghai Ind. Park2026-02-25

7.2 Sample Test Results

No.Val 1(mg/kg)Val 2(mg/kg)Val 3(mg/kg)Mean(mg/kg)RSD(%)
SW-2026-00125.3525.4825.4225.420.47
SW-2026-00212.1812.2512.2212.220.49
SW-2026-0033.853.923.883.881.03

7.3 Sample Spike Test

No.Spike Conc(mg/kg)Spiked Val(mg/kg)Rec(%)
SW-2026-00140.065.58100.4
SW-2026-00240.052.35100.3
SW-2026-00340.043.92100.1

Conclusion: Results valid. Rec 100.1%~100.4%. Method fits sample analysis

8. Full Sample Prep Validation

8.1 Sample Prep Procedure

Per HJ 687-2014, full validation including sample preparation:

Sampling & storage:Stainless sampler, sealed, 4℃, test within 7d

Pretreatment:Air-dry, grind, sieve 0.15mm, mix

Alkali digestion: 2.5g sample + 50mL 0.1 mol/L NaOH, boil bath 30min

Centrifuge: 4000 r/min, 10min

Supernatant: Direct injection

8.2 Validation Results

ItemContentResultConclusion
Sample RepresentHomogeneity test RPD<3%Pass
Extraction EffSpike Rec99.8%~101.5%Pass
Operation NormSOPcomplu with HJ 687-2014Pass
Data IntegrityRaw record traceableCompletePass
Min Sample Mass≥2.5g2.5gComply

9. Validation Conclusion

9.1 Overall Assessment

ItemResultMeet HJ 687-2014
InstrumentParameters reasonable,performance okYes
Cal Curver = 0.9999Yes(r≥0.995)
MDL1.64 mg/kgBetter than std(2 mg/kg)
LOQ6.56 mg/kgBetter than std(8 mg/kg)
PrecisionRSD = 0.03%~0.59%Yes(≤5%~10%)
AccuracyRec99.8%~101.5%Yes(85%~115%)
Real SampleValid test, Rec100.1%~100.4%Yes
Sample PrepFull procedure standardizedYes

9.2 Final Conclusion

Validated method. AA-1800C AAS for Cr(VI) in solid waste per HJ 687-2014. MDL, LOQ, precision & accuracy meet std. Full sample prep standardized, data traceable. Passed, for routine test.

Document Information

Report No.: MACY-2026-0305
Document Code: 20260305-AA1800CN
Date: March 2026
Produced by: Shanghai Macylab Instrument Inc.

2026 Shanghai Macylab Instrument Inc. All rights reserved.

The information provided in this document is for reference purposes. Shanghai Macylab Instrument Inc. makes reasonable efforts to ensure the accuracy of the information but shall not be liable for errors or for any indirect losses arising from the use of this document.

Specifications, instructions, and technical information are subject to change without prior notice.

Official Website: www.macylab.com