Method Validation Report for Cr(Ⅵ) in Soil & Sediment

1.Validation Overview

Item:Method Validation Report for Cr(Ⅵ) in Soil & Sediment

REF: HJ 1082—2019Instrument:AA-1800E Atomic Absorption Spectrometer

2.Instrument Conditions

2.1 Main Parameters

Wavelength: 357.9 nm

Bandwidth: 0.4 nm

Lamp current: 6 mA

Burner height: 8 mm

Burner: 50 mm Ti head

Acetylene: 1.8 L/min

Air: 8.5 L/min

Integration: 3 s

Mode: Peak height

Replicates: 3, averaged

2.2 Instrument Performance

ItemSpecActualConclusion
Static Baseline Drift≤±0.002A/30min±0.0012A/30mPass
Dynamic Baseline Drif≤±0.005A/30min±0.0028A/30minPass
Precision(RSD)≤0.5%0.32%Pass
Wavelength Accuracy≤0.15nm0.08nmPass

3.Calibration Curve Validation

3.1 Standard Solution Prep

Certified 1000 mg/L Cr(Ⅵ) stock, serially diluted by 0.1 mol/L NaOH:

No.Conc.(mg/L)Method
00.00Blank(0.1 mol/L NaOH)
10.50Serial dilution
21.00Serial dilution
32.00Serial dilution
44.00Serial dilution
56.00Serial dilution
68.00Serial dilution

3.2 Calibration Curve Data

Conc.(mg/L)A1A2A3Average
0.000.00120.00100.00110.0011
0.500.08250.08310.08280.0828
1.000.16520.16480.16550.1652
2.000.32980.33050.32920.3300
4.000.65850.65920.65780.6585
6.000.98520.98450.98580.9852
8.001.31251.31181.31321.3125

3.3 Curve Fitting Result

Regression:y = 0.1641x + 0.0008

Correlation coefficient:(r)0.9998

Intercept:0.0008

Slope:0.1641

Criterion:r ≥ 0.999Conclusion:r = 0.9998,meets requirement

4. MDL & LOQ Verification

4.1 Method Detection Limit (MDL) Test

Per HJ 1082-2019, with 5.0 g soil sample and final volume 100 mL, MDL = 0.5 mg/kg .

Test procedure:

Prepare standard solution near detection limit(0.1 mg/L)

Measure for 7 consecutive runs

Calculate MDL = t × S (t=3.143, df=6, 99% confidence)

No.AbsorbanceConc.(mg/L)
10.01650.100
20.01580.096
30.01720.104
40.01610.098
50.01680.102
60.01550.094
70.01630.099

Calculation:

Mean:0.099 mg/L

S:0.0035 mg/L

MDL = 3.143 × 0.0035 = 0.011 mg/L(soln)

Soil MDL:0.011 × 100 / 5.0 = 0.22 mg/kg

4.2 LOQ Verification

ItemSpecResultConclusion
MDL0.5 mg/kg0.22 mg/kgExceeds spec
LOQ2.0 mg/kg0.88 mg/kgExceeds spec

Rule: LOQ=4×MDL

Verdict: MDL & LOQ comply with HJ10822019

5. Precision Verification

5.1 Scheme

HJ10822019: low/med/high levels, 6 parallels each

LevelSpike(mg/kg)Soln(mg/L)
Low2.00.10
Med10.00.50
High40.02.00

5.2 Precision Test Results

Low level(2.0 mg/kg)

No.Result(mg/kg)
12.08
22.12
32.05
42.10
52.07
62.09

Mean:2.08 mg/kg

Std Dev:0.029 mg/kg

RSD = 1.40%Medium level(10.0 mg/kg)

No.Result(mg/kg)
110.15
210.22
310.08
410.18
510.12
610.20

Mean:10.15 mg/kg

Std Dev:0.057 mg/kg

RSD = 0.56%

High level(40.0 mg/kg)

No.Result(mg/kg)
140.25
240.18
340.32
440.20
540.28
640.15

Mean:40.22 mg/kg

Std Dev.:0.065 mg/kg

RSD = 0.16%

5.3 Precision Conclusion

LevelSpike(mg/kg)RSD(%)SpecResult
Low2.01.40≤10%Pass
Med10.00.56≤5%Pass
High40.00.16≤5%Pass

Rule: HJ10822019: RSD ≤10%(low), ≤5%(med/high) Verdict: All three levels comply, precision qualified.

6. Accuracy Verification

6.1 Scheme

Accuracy was verified using certified soil reference material via spike recovery test.

Reference material information:

Name: Certified Reference Material for Hexavalent Chromium in Soil

CRM:RMH-A048

Certified Cr(VI):28.8 mg/kg

Uncertainty:±2.5 mg/kg

6.2 Spike Recovery

Low spike(2.0 mg/kg)

No.Background(mg/kg)Spiked(mg/kg)Recovery(%)
10.122.18103.0
20.152.22102.7
30.102.15103.5
40.132.20103.1
50.112.17103.3
60.142.21102.9

Avg recovery:103.1%

Recovery range:102.7% ~ 103.5%

Medium spike(10.0 mg/kg)

No.Background(mg/kg)Spiked(mg/kg)Recovery(%)
10.1510.28101.3
20.1210.22101.0
30.1810.35101.7
40.1410.25101.1
50.1610.30101.4
60.1310.23101.0

Avg recovery:101.3%

Range:101.0% ~ 101.7%

High spike(40.0 mg/kg)

No.Background(mg/kg)Spiked(mg/kg)Recovery(%)
10.1840.3599.9
20.1540.2899.7
30.2040.42100.0
40.1740.3299.8
50.1940.3899.9
60.1640.2599.6

Avg recovery:99.8%

Range:99.6% ~ 100.0%

6.3 CRM Verification

Soil CRM RMHA048 (certified value 28.8 mg/kg) was tested:

No.Measured(mg/kg)
128.52
228.68
328.75
428.62
528.70
628.65

Mean:28.65 mg/kg

Certified value:28.8 mg/kg

Relative error:(28.65 – 28.8) / 28.8 × 100% = -0.52%

RSD:0.28%

Criterion:CRM measured value shall lie within certified uncertainty interval.Verdict: 28.65 mg/kg falls in 28.8 ± 2.5 mg/kg, compliant

6.4 Accuracy Conclusion

LevelSpike(mg/kg)Avg Recovery(%)SpecResult
Low2.0103.185%~115%Pass
Med10.0101.390%~110%Pass
High40.099.890%~110%Pass
CRM28.8RE-0.52%≤10%Pass

Verdict: Recoveries of three levels and CRM test satisfy criteria, method accuracy qualified.

7. Actual Sample Verification

7.1 Sample Information

Sample IDTypeLocationDate
SP-2026-001SoilIndustrial park,Shanghai2026-06-28
SP-2026-002SedimentRiver channel, Shanghai2026-06-29
SP-2026-003SoilFarmland, Shanghai2026-06-30

7.2 Test Results

Sample IDtest 1(mg/kg)test 2(mg/kg)test 3(mg/kg)AVG(mg/kg)RSD(%)
SP-2026-0013.253.323.283.281.08
SP-2026-0021.851.921.881.881.89
SP-2026-0030.420.450.430.433.45

7.3 Actual Sample Spike Test

Sample IDSpike(mg/kg)Measured(mg/kg)Recovery(%)
SP-2026-00110.013.35100.7
SP-2026-00210.011.92100.5
SP-2026-00310.010.48100.5

Conclusion: Actual sample results are reasonable; spike recoveries 95%–105%. This method is suitable for real sample analysis

8.Whole-process Sample Preparation Verification

8.1 Sample Preparation Process

In accordance with the requirements of HJ 1082-2019, the validation covers the entire process, including sample preparation:

Collection & storage: Stainless steel sampler, 4℃ preservation, test within 24 h

Pretreatment: Air dry, grind, 0.15 mm sieve

Alkaline extraction: 5.0 g sample + 50 mL 0.1 mol/L NaOH, shake 30 min

Centrifuge: 4000 r/min, 10 min

Test: Direct injection of supernatant

8.2 Verification Results

ItemContentResultConclusion
Sample RepresentativenessHomogeneity checkRD<5%Pass
Extraction EfficiencySpike recovery99.8%~103.1%Pass
Operation ComplianceSOP executionMeetHJ 1082-2019Pass
Data IntegrityTraceable raw dataCompletePass

9. Verification Conclusion

9.1 Comprehensive Assessment

ItemResultCompliance HJ 1082-2019
InstrumentParameters reasonable,performance qualifiedCompliant
Calibration Curver = 0.9998Compliant(r≥0.999)
MDL0.22 mg/kgBetter than(0.5 mg/kg)
Quant Limit0.88 mg/kgBetter than(2.0 mg/kg)
PrecisionRSD = 0.16%~1.40%Compliant(≤5%~10%)
AccuracyRecovery 99.8%~103.1%Compliant(85%~115%)
Real SamplesValid results,recovery100.5%~100.7%Compliant
Sample PrepFull process standardizedCompliant

9.2 Final Conclusion

Validated with AA1800E AAS per HJ 10822019 for Cr(VI) in soil & sediment. All performance indicators satisfy requirements. Sample preparation is standardized, data traceable. Method is qualified for routine analysis.

10. Document Information

Validation Organization: Shanghai Macylab Instrument Inc.

Report No.: MACY-2026-0705

Validation Date: July 05, 2026

This application validation report is applicable to the sample and test conditions described in this document. Reproduction or distribution of this report, in whole or in part, requires prior written authorization from Shanghai Macylab Instrument Inc.

2026 Shanghai Macylab Instrument Inc. All rights reserved.