Method Validation Report for the Determination of 11 Elements in Soil and Sediments

1.Verification Overview

Verification Purpose

To confirm the applicability of the method for determining 11 elements in soil and sediment using alkali fusion–inductively coupled plasma optical emission spectrometry (ICP-OES) under laboratory conditions, a systematic validation was conducted covering instrument conditions, calibration curves, detection limits, lower limit of quantification, precision, accuracy, and actual sample analysis, ensuring that the method complies with the requirements of the HJ 974-2018 standard.

2.Verification Basis

1.Determination of 11 elements in soil and sediment – Alkaline fusion–inductively coupled plasma optical emission spectrometry – HJ 974-2018

2.Verification Regulation of Emission Spectrometer JJG 768

3.General Principles for the Validation of Chemical Analysis Methods GB/T 27404

4.Technical Specification for Reference Materials for Soil Composition Analysis GB/T 15000

3.Instrumentation and Reagents

3.1 Primary instruments and equipment

Inductively coupled plasma optical emission spectrometer ICP-6810-A2 Shanghai Macylab Instrument Inc.

High-temperature muffle furnace

Platinum crucible 50mL

electronic balance 0.1mg Mettler Toledo Microwave Digestor

3.2 Instrument Main Parameters

Light source: Solid-state RF generator, frequency 27.12 MHz;

Burner: Vertical burner with cold-cone flamelet elimination technology;

Detector: High-speed surface-array CCD; Single-exposure full-spectrum direct

reading; Atomization

system: Mass flow controller for regulating carrier gas, auxiliary gas, and cooling

gas; Linear

range: Wide dynamic linear range, suitable for trace-to-constant concentration analysis.

4. Verification Element and Wavelength Selection Switch

Verify 11 elements per HJ 974-2018:

No.ElementWacelength(nm)Interference check
1Mn257.610No Interference
2Ba455.403No Interference
3V292.402No Interference
4Sr460.733No Interference
5Ti334.941No Interference
6Ca317.933No Interference
7Mg285.213No Interference
8Fe238.204No Interference
9Al396.152No Interference
10K766.491No Interference
11Si251.611No Interference

5. Verification Items and Results

5.1 Instrumental Condition Optimization Method

ParameterSetpointRemark
Radio-frequency power1300 WSample matrix optimization
Nebulizer gas flow0.65 L/minMFC control
Aux gas flow1.20 L/minPlasma stable
Cool gas flow15.00 L/minProtect torch
Obs height15 mmVertical torch opt
Int time3 sGuarantee SNR
Replicates3 timesTake avg
Fusion temp1000℃Na₂CO₃-Li₂O flux
Fusion time30 minFull fusion

5.2 Calibration curve verify

Std soln prep: Multi-element mixed std from CRM, conc gradients: 0、1.0、5.0、10.0、20.0、50.0 mg/L.

ElemWavelength(nm)Range(mg/L)RIntercept
Mn257.6101.0~50.00.99980.0015
Ba455.4031.0~50.00.99970.0012
V292.4021.0~50.00.99960.0018
Sr460.7331.0~50.00.99970.0014
Ti334.9411.0~50.00.99950.0020
Ca317.9331.0~50.00.99980.0011
Mg285.2131.0~50.00.99970.0013
Fe238.2041.0~50.00.99980.0010
Al396.1521.0~50.00.99960.0016
K766.4911.0~50.00.99970.0012
Si251.6111.0~50.00.99950.0019

Note: All r ≥0.9995, meets HJ 974-2018.

5.3 MDL & LOQ

Per HJ 974-2018 App A: 0.2 g sample, 50 mL final volume. MDL=3×blank SD, LOQ=10×blank SD.

ElemMDL(mg/L)LOQ(mg/L)MDL(mg/kg)LOQ(mg/kg)
Mn0.050.170.0130.043
Ba0.080.270.0200.068
V0.030.100.0080.025
Sr0.060.200.0150.050
Ti0.100.330.0250.083
Ca0.150.500.0380.125
Mg0.080.270.0200.068
Fe0.120.400.0300.100
Al0.100.330.0250.083
K0.100.330.0250.083
Si0.150.500.0380.125

Conclusion: MDL & LOQ of all elements meet HJ 974-2018.

5.4 Precision Test (LowMedHigh levels)

Test sample: Soil CRM GSS-39. Spiked 3 levels, 6 replicates per level.

LevelElemSpike(mg/kg)Measured(mg/kg)Average(mg/kg)SDRSD(%)
LowMn5049.2, 50.1, 49.8, 50.5, 49.5, 50.249.880.420.84
LowBa3029.5, 30.2, 29.8, 30.5, 29.3, 30.129.930.411.37
LowV2019.8, 20.3, 19.9, 20.5, 19.6, 20.220.070.321.59
MidMn200198.5, 201.2, 199.8, 200.5, 198.2, 200.1199.581.250.63
MidBa120118.5, 121.2, 119.8, 120.5, 118.2, 120.1119.581.251.05
MidV8079.5, 81.2, 79.8, 80.5, 79.2, 80.179.950.851.06
HighMn500498.2, 501.5, 499.8, 500.5, 498.5, 500.2499.581.350.27
HighBa300298.5, 301.2, 299.8, 300.5, 298.2, 300.1299.581.350.45
HighV200198.5, 201.2, 199.8, 200.5, 198.2, 200.1199.581.350.68

Precision criteria: RSD ≤5% (all levels)

Conclusion: All RSD ≤1.59%, compliant, good precision.

5.5 Trueness Test (Spike recovery)

Test sample: Soil CRM GSS-39, spike-recovery method.

LevelElemCertified(mg/kg)Founde(mg/kg)Rec(%)Avg Rec.(%)
LowMn48.548.299.499.2
LowBa28.528.399.399.1
LowV18.518.499.599.3
MidMn195.0193.899.499.3
MidBa115.0114.299.399.2
MidV75.074.599.399.2
HighMn485.0482.599.599.4
HighBa285.0283.299.499.3
HighV185.0183.899.499.3

Criteria: Recovery: 95%-105%

Conclusion: Recoveries 99.1%-99.5% for all levels. Good trueness; measured values consistent with certified values.

5.6 Actual sample test

Sample source: Soil samples from different sites

Sample IDElemResult (mg/kg)Note
Soil-001Mn125.550ml final
Ba85.2
V45.8
Sr125.3
Ti285.6
Ca1250.5
Mg385.2
Fe2850.5
Al15800.5
K18500.5
Si28500.5
Soil-002Mn98.550ml final
Ba72.3
V38.5
Sr105.8
Ti245.2
Ca1120.8
Mg325.5
Fe2580.2
Al14200.5
K16800.2
Si26500.8

Conclusion: Real-sample results reasonable. Method suitable for 11-element test in soil & sediment.

6. Wholeprocess validation of sample preparation

6.1 Sample pretreatment procedure

StepOperationKey control points
1Sample collection & storagePE bag, avoid contamination
2Sample air-dryingRoom temp, dark, no high temp
3Grinding & sieving100-mesh nylon sieve, no metal pollution
4Weighing0.2 g (accurate to 0.1 mg)
5Alkali fusion2 g Na₂CO₃-Li₂O (1:1) flux, 1000℃, 30 min
6LeachingHot-water leach melt, 50mL final
7Instrument testICP-OES measurement

6.2 QC for alkalifusion process

QC itemRequirementExecution status
Flux purityGR gradeGR reagents applied
Pt crucibleIntact & cleanChecked,good
Fusion temp1000±20℃Calibrated, compliant
Fusion time30±2minStrictly controlled
Leaching waterUPW(18.2MΩ·cm)Adopted

7.Interference & Elimination

7.1 Per HJ 9742018, alkalifusion ICPOES interferences:

Interf. TypeSourceElimination
SpectralBkg&line overlapBkg corr.,interf.coeff.
MatrixHigh-salt matrixMatrix match, dilution
ChemicalRefractory compoundsFull alkali-fusion
PhysicalVisc.&surface-tensionInt.std,matrix match

Lab measures:

Complete soil matrix decomposition via alkali fusion

Matrix-matched standard solutions

Select interference-free / low-interference wavelengths

Cold cone tail-flame removal to reduce self-absorption

8. QA & QC

QC ItemReq.Status
Blank testBlank<MDLDone, ok
DuplicateRD≤10%Done, ok
CRM testIn certified rangeDone, ok
Spike rec.95%~105%Done, ok
Calib. curveR≥0.999Done,r≥0.9995
Fusion completenessDissolved, no residueChecked, ok

9. Verify conclusions

Systematically validated: Alkali fusion–inductively coupled plasma optical emission spectrometry(ICP-6810-A2)Method for determining 11 elements in soil and sediment:

Valid. ItemResultCompl.
InstrumentReasonable params, stable✓ Yes
Calib. curver≥0.9995✓ Yes
MDLHJ 974-2018 Appendix A✓ Yes
LOQHJ 974-2018 Appendix A✓ Yes
PrecisionRSD≤1.59%(all levels)✓ Yes
AccuracyRec.99.1%~99.5%✓ Yes
Real sampleValid, applicable✓ Yes
Sample prepFull fusion,standard✓ Yes

Overall conclusion: This method demonstrates stable performance under laboratory conditions, and all technical parameters comply with the requirements of the HJ 974-2018 standard; it has been successfully validated and is therefore suitable for formal application in the determination of 11 elements—manganese, barium, vanadium, strontium, titanium, calcium, magnesium, iron, aluminum, potassium, and silicon—in soils and sediments.

10. Document Information

Validation Organization: Shanghai Macylab Instrument Inc.

Report Number: MACY-2026-0626

Document Code: 20260626-ICP6810CN

Validation Date: June 26, 2026

This application validation report is applicable to the sample and test conditions described in this document. Reproduction or distribution of this report, in whole or in part, requires prior written authorization from Shanghai Macylab Instrument Inc.