Key Features

  • Double Beam Optical System: Splits the light path to simultaneously measure the sample and reference. This design actively compensates for light source fluctuations, improving baseline stability (0.0003A/h @500nm) and minimizing background noise for long-term measurements.
  • Adjustable Spectral Bandwidth: Features five selectable bandwidth options (0.5, 1.0, 2.0, 4.0, and 5.0 nm). This accommodates specific resolution requirements for various testing standards and chemical analyses.
  • 10-Inch Touch LCD Screen: Allows users to configure testing parameters, establish standard curves, and view analytical graphs directly on the large display without requiring a connected computer.
  • Comprehensive Software Functions: The onboard software supports quantitative measurement, kinetic testing, multi-wavelength determination, full-band or segmented wavelength scanning, and specific DNA/protein purity analysis.
  • Flexible Data Management & PC Control: Retains measurement parameters and test data even after powering down. It features a USB interface for outputting measurement data directly to a printer and supports optional PC control for expanded data manipulation.

Specifications

ParameterSpecification
Optical Path StructureDouble beam spectrophotometer
Wavelength Range190 ~ 1100 nm
Wavelength SettingAutomatic
Band Width0.5 / 1.0 / 2.0 / 4.0 / 5.0 nm (Adjustable)
Wavelength Accuracy±0.3 nm
Wavelength Repeatability0.1 nm
Photometric Display Range-4 ~ 4 A
Photometric Accuracy±0.2% T
Photometric Repeatability0.1% T
Stray Light≤0.02% T
Stability0.0003 A/h @ 500nm
Photometric ModeT, A, C, E
Display ModeTouch LCD Screen (10 inches)
DetectorImported Silicon Photodiode
Light SourceImported Deuterium Lamp & Tungsten Halogen Lamp
Data OutputUSB interface for printer
Power RequirementAC 220V/50Hz or 110V/60Hz
Dimensions (W×D×H)650 × 430 × 240 mm
Weight21 Kg

Standard Configuration

mv 2300

Applications

The M-2800S is utilized for photometric testing across the ultraviolet and visible light spectrums:

  • Life Sciences & Biology: Direct quantification of DNA, RNA, and protein concentrations, as well as specific nucleic acid purity testing using the onboard software.
  • Pharmaceutical Analysis: Quality control of active pharmaceutical ingredients (APIs), standard multi-wavelength quantitative tests, and multi-order derivative determinations.
  • Environmental Protection: Testing water quality parameters and trace compounds that exhibit absorption peaks in the UV range.
  • Material Science & Chemical Research: Spectral scanning to identify chemical structures and monitor kinetic reactions over specific time intervals.

Who is it for?

  • Research Universities and Institutes: Laboratories requiring a double-beam UV-Vis spectrophotometer with adjustable bandwidths for diverse analytical research and teaching applications.
  • Pharmaceutical QC Departments: Quality control labs demanding precise optical performance, low stray light, and reliable standard curve generation for compliance testing.
  • Environmental Monitoring Stations: Facilities that analyze water and soil extracts, benefiting from the instrument’s stable baseline and broad wavelength range.

Why choose it?

  • Adaptable Optical Resolution: The ability to switch between five different spectral bandwidths (from 0.5nm up to 5.0nm) provides the flexibility needed to test both broad peaks and fine spectral details on a single instrument.
  • Enhanced Baseline Stability: The double-beam optical structure provides high stability for kinetic studies and prolonged analytical tests, maintaining accuracy over time.
  • Certified Qualifications: The instrument complies with major international product qualifications and safety standards, holding certifications including the European CE mark, North American ETL, FCC, RoHS, China CQC, and 3C.

Software features

DNA Protein

ul 1000
ul 1000

Kinetics

ul 1000
ul 1000

Multi Wavelength

ul 1000
ul 1000

Quantitive

ul 1000
ul 1000

WL Scan

ul 1000
ul 1000

Packaging, Installation & Operation

Packaging & Shipping

After unpacking the instrument, compare the contents with the packing list and check that all items are present and undamaged.

Installation & Setup

Place the spectrophotometer in a suitable location away from direct sunlight, dust, corrosive gases, and strong vibrations. For stable operation, keep it away from strong magnetic or electrical fields and electrical devices that may generate high-frequency fields.

Keep the area around and underneath the instrument clear to allow adequate airflow.

Use the appropriate power cord and connect the instrument to a grounded outlet.

Initial Operation

Turn on the spectrophotometer and allow it to warm up for the recommended time before taking measurements. Follow the instrument’s operating instructions to complete any required system initialization, calibration, or wavelength verification before measurement.

products-Packaging-Shipping.

Exhibition

products-Exhibition.

FAQ (Frequently Asked Questions)

Yes. The M-2800S operates in the UV range and features built-in software specifically for DNA and protein tests, allowing for the direct analysis of nucleic acid purity and protein concentration.

No. The instrument is equipped with a 10-inch touch LCD screen and onboard software that supports wavelength scanning, kinetics, and quantitative analysis. However, it can also be operated via PC control if your workflow requires it.

The M-2800S features an adjustable spectral bandwidth with five settings: 0.5, 1.0, 2.0, 4.0, and 5.0 nm.

Yes. The device is designed to save measurement parameters and test data when powered down, preventing the loss of experimental setups.

The instrument utilizes a combination of an imported deuterium lamp for the ultraviolet range (190-340nm) and a tungsten halogen lamp for the visible range (340-1100nm).

×