The Macylab M-2900 is a highly accurate double-beam UV-Vis spectrophotometer operating across the 190nm to 1100nm wavelength range. Engineered for demanding analytical requirements, it allows operators to choose from six adjustable spectral bandwidths (0.1, 0.2, 0.5, 1.0, 2.0, and 4.0 nm) to optimize optical resolution based on the sample type.
Equipped with a 10-inch touch LCD screen and an integrated operating system, the instrument supports complex photometric analyses directly on the device. It utilizes an imported deuterium and tungsten halogen lamp combination alongside a silicon photodiode detector, making it a reliable platform for biological research, pharmaceuticals, and environmental protection.
Key Features
Double Beam Optical System: Splits the light path to simultaneously measure the sample and reference, significantly improving baseline stability (0.0003A/h @500nm) and minimizing background noise for long-term measurements.
Adjustable Spectral Bandwidth: Features six selectable bandwidth options (0.1, 0.2, 0.5, 1.0, 2.0, and 4.0 nm). This accommodates strict resolution requirements for various pharmacopeia standards and specific chemical analyses.
10-Inch Touch LCD Screen: Allows users to configure testing parameters, establish quantitative standard curves, and view analytical graphs directly on the large display without needing a connected computer.
Comprehensive Software Functions: The onboard software supports standard quantitative measurement, kinetics, multi-wavelength testing, full-band or segmented wavelength scanning, and specific DNA/protein purity analysis.
Flexible Data Management: Retains measurement parameters and test data after power down. It features a USB interface for outputting measurement data to a printer, and supports optional PC control for extended data manipulation.
Specifications
Parameter
Specification
Optical Path Structure
Double beam spectrophotometer
Wavelength Range
190 ~ 1100 nm
Wavelength Setting
Automatic
Band Width
0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 4.0 nm
Wavelength Accuracy
±0.3 nm
Wavelength Repeatability
0.1 nm
Photometric Display Range
-4 ~ 4 A
Photometric Accuracy
±0.2% τ
Photometric Repeatability
0.1% τ
Stray Light
≤0.02% τ
Stability
0.0003 A/h @ 500nm
Photometric Mode
T, A, C, E
Display Mode
Touch LCD Screen (10 inches)
Detector
Imported Silicon Photodiode
Light Source
Imported Deuterium Lamp & Tungsten Halogen Lamp
Data Output
USB interface for printer
Power Requirement
AC 220V/50Hz or 110V/60Hz
Dimensions (W×D×H)
650 × 430 × 230 mm
Weight
21 Kg
Standard Configuration
Applications
The M-2900 is designed for high-resolution photometric testing across the ultraviolet and visible light spectrums:
Life Sciences & Biology: Direct quantification of DNA, RNA, and protein concentrations, as well as specific nucleic acid purity testing.
Pharmaceutical Analysis: Quality control of active pharmaceutical ingredients (APIs), standard multi-wavelength quantitative tests, and multi-order derivative determinations.
Environmental Protection: Testing water quality parameters, including trace heavy metals and organic compounds that exhibit absorption peaks in the UV range.
Material Science & Chemical Research: High-resolution spectral scanning to identify chemical structures and monitor kinetic reactions over specific time intervals.
Who is it for?
Research Universities and Institutes: Laboratories requiring a high-resolution, double-beam UV-Vis spectrophotometer with adjustable bandwidths for diverse and complex analytical research.
Pharmaceutical QC Departments: Quality control labs demanding precise optical performance, low stray light, and reliable standard curve generation for strict compliance testing.
Environmental Monitoring Stations: Facilities that analyze complex water and soil extracts, benefiting from the instrument’s stable baseline and broad wavelength range.
Why choose it?
Adaptable Optical Resolution: The ability to switch between six different spectral bandwidths (from 0.1nm up to 4.0nm) provides the flexibility needed to test both broad peaks and fine spectral details on a single instrument.
Enhanced Baseline Stability: The double-beam optical structure actively compensates for light source fluctuations, resulting in a highly stable baseline that is critical for kinetic studies and prolonged analytical tests.
Certified Qualifications: The instrument complies with major international product qualifications and safety standards, holding certifications including the European CE mark, North American ETL, FCC, RoHS, China CQC, and 3C.
Software features
DNA Protein
Kinetics
Multi Wavelength
Quantitive
WL Scan
Packaging, Installation & Operation
Packaging & Shipping
After unpacking the instrument, compare the contents with the packing list and check that all items are present and undamaged.
Installation & Setup
Place the spectrophotometer in a suitable location away from direct sunlight, dust, corrosive gases, and strong vibrations. For stable operation, keep it away from strong magnetic or electrical fields and electrical devices that may generate high-frequency fields.
Keep the area around and underneath the instrument clear to allow adequate airflow.
Use the appropriate power cord and connect the instrument to a grounded outlet.
Initial Operation
Turn on the spectrophotometer and allow it to warm up for the recommended time before taking measurements. Follow the instrument’s operating instructions to complete any required system initialization, calibration, or wavelength verification before measurement.
Exhibition
FAQ (Frequently Asked Questions)
Yes. The M-2900 covers the full UV-Vis spectrum from 190nm to 1100nm, utilizing a deuterium lamp for the ultraviolet range and a tungsten halogen lamp for the visible range.
No. The instrument is equipped with a 10-inch touch LCD screen and onboard software that supports wavelength scanning, kinetics, quantitative analysis, and DNA/protein tests. However, it can also be operated via PC control if your workflow requires it.
The M-2900 features an adjustable spectral bandwidth with six settings: 0.1, 0.2, 0.5, 1.0, 2.0, and 4.0 nm. This allows you to select the appropriate optical resolution for your specific sample requirements.
Yes. The device is designed to save measurement parameters and test data when powered down, preventing the loss of experimental setups.
It utilizes a double beam spectrophotometer structure, which splits the light into a sample beam and a reference beam to improve photometric accuracy and baseline stability.