ICP-6800E is a full-spectrum direct-reading inductively coupled plasma emission
spectrometer for the determination of trace and trace elements in different substances (dissolved in nitric acid, hydrochloric acid, hydrofluoric acid, etc.).
Performance characteristics
ICP-6800E is professionally designed for the analysis and determination of rare earth elements, lanthanide elements La, Ce, Pr, Nd, Pm, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb, Lu, and two elements closely related to 15 elements of lanthanide system – Sc and Y total 17 elements, can be used for the analysis of the composition of rare earth ores; the examination of raw materials and finished products of rare earth products.
RF Power technical parameter
Circuit type: solid-state RF power supply, with function of automatch
Frequency: 27.12MHz±0.05%
Frequency Stability: <0.1%
Power Output: 500—1600W
Power Output Stability: <0.3%
Escaped RF radiation: 30cm away from the instrument, electric field: E﹤2V/m
Sampling System Technical Parameter
Output working coil inner diameter :25mm
Torque tube: Three concentric, external diameter 20mm
Cooling Water:Temperature: 20-25℃, Rate of Flow>5L/min, Hydraulic Pressure>0.1MPa
Spectrometer
Optics:Czerny-Turner type
Focal length: 1000 mm
Grating: Ion Beam Etching Holographic Grating, 3600L/mm or 2400L/mm
Reciprocal linear dispersion: 0.26nm/mm
Resolution:≤ 0.007nm (3600 line grating);≤0.015nm (2400line grating)
Wavelength range: 3600 line grating:(190nm~500) nm;2400 line grating: (190nm~800) nm
Minimum pace of stepping motor: ≤0.0006 nm
Exit Slit:12μm;Entrance Slit:10μm
Photoelectric Converter Performance
Photomultiplier tube specification: R293/R928
Negative HV on PMT: 0-1000V;Stability: <0.05%
Detection Limits (μg/L)
Element
WL(nm)
Limit
Element
WL(nm)
Limit
La
408.672
<3.0
Cr
267.716
<5.0
Ce
413.765
<5.0
Al
396.152
<5.0
Pr
414.311
<5.0
Zr
343.823
<5.0
Nd
401.225
<5.0
Ag
328.068
<3.0
Sm
360.946
<10.0
Sr
407.771
<1.0
Eu
381.967
<1.0
Au
242.795
<5.0
Gd
342.247
<10.0
Pt
265.945
<5.0
Tb
350.917
<3.0
Pd
340.458
<5.0
Dy
353.170
<3.0
Ir
224.268
<10.0
Ho
345.600
<3.0
Rh
343.489
<10.0
Er
337.271
<3.0
Ru
240.272
<5.0
Tm
313.126
<3.0
Ba
455.403
<1.0
Yb
369.419
<1.0
As
228.812
≤15
Lu
261.541
<3.0
Sb
206.833
≤15
Y
371.030
<1.0
Bi
223.061
≤10
Sc
335.373
<1.0
Hg
253.652
≤15
Ta
226.230
<5.0
Pb
220.353
≤15
Nb
313.340
<5.0
Ga
294.364
≤10
Mn
257.610
<3.0
Se
203.985
≤10
Mg
279.553
<1.0
Sn
242.949
≤20
B
249.773
<10.0
Te
214.281
≤10
Zn
213.856
<3.0
Ta
226.230
≤5.0
Co
228.616
<3.0
Th
283.730
≤10
Si
251.611
<10.0
Tl
276.787
≤30
Ni
232.003
<5.0
Re
227.525
≤5
Cd
226.502
<3.0
Ge
209.426
≤15
Fe
239.562
<3.0
Os
225.585
≤1
Ca
393.366
<1.0
W
207.911
≤10
Mo
281.615
<5.0
Cu
324.754
<3.0
V
310.230
<5.0
Li
670.784
≤3
Be
313.041
<1.0
Na
588.995
≤20
Ti
334.941
<3.0
K
766.490
≤60
Certifications
Product Qualification: European safety qualification mark CE, North American safety certification ETL, Federal Communications Commission FCC, European electrical safety certification RoHS, China Quality Certification Center CQC, China national mandatory product certification 3C
Packaging & Shipping
After carefully unpacking the contents, check the materials with the packing list (page 4) to ensure that you have received everything in good condition.
Place the instrument in a suitable location away from direct sunlight. In order to have the best performance from your instrument, keep it as far as possible from any strong magnetic or electrical fields or any electrical device that may generate high-frequency fields. Set the unit up in an area that is free of dust, corrosive gases and strong vibrations.
Remove any obstructions or materials that could hinder the flow of air under and around the instrument.
Use the appropriate power cord and plug into a grounded outlet.
Turn on your spectrophotometer. Allow it to warm up for 15 minutes before taking any readings. We suggest you then do the Calibrate System with the Search 656.1nm to set the wavelength to the deuterium lamp emission line.
Exhibition
FAQ
Q: How to pick up the best among these models?
A:Tell us your request on product and almost target price,we recommend you the most suitable model. Selling you the most expensive product is not our goal,selling you the most suitable is!
Q: How to place order or pay?
A:You can place order to us directly,we offer you PI (contract) with bank information, you pay here,we ship the product,Every process is based on PI.
Q: How to transport the product?
A:Depending on the quantity and what kind of machine you order or whether you are urgent or not to check how big it is to check the freight. By sea: More than 60kg or big quantity;Safe,cheap but slow; Usually 11-30 days to North America; 30-45 days to Middle of South America; 40-60 days to South America; 15-28 days to South Africa; 25-33 days to Europe; 30-45 days to East Europe; 35-45 days to North Europe; 28-35 days to Mediterranean countries; 28-40 days to Middle East countries; 7-20 days to Southeast Asia countries; By air to airport: Suit for big product or quantity but urgent a lot. Usually 7-10 days arrive. Some direct flight can arrive at 3-5 days. By Express delivery such as DHL,FedEx,TNT: Suit for small or not too heavy and very urgent product. Generally 5-7 days arrive.
Q: What kind of warranty will be offer?
A:Most of our product we offer 1 years warranty,except perishable part. During the warranty time,any problem,we offer after-sale service; Welcome our customer talk this with our sales directly,they will offer you more details;
Q: How’s our service?
A: Before place order,any inquiry we make sure will reply in 1 day,and offer you our suitable advice based on our professional product knowledge; After place order,we will help you follow the shipping information until you get these machines; If any problems on delivery or using,we make sure you can contact us in 1 day.