The Macylab ICP-6810-A2 is a full-spectrum, direct-reading Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) engineered for the quantitative determination of trace and major elements. Operating across a broad wavelength range of 165nm to 950nm, the instrument analyzes samples digested in various acids, including nitric acid, hydrochloric acid, and hydrofluoric acid.
The system utilizes an echelle grating and prism cross-dispersion optical design with no moving optical parts. Combined with a large-area Charge-Coupled Device (CCD) detector, the ICP-6810-A2 captures the intensity integral values of all analyzed spectral lines in a single exposure, allowing for the complete measurement of elements in a sample within one minute.
Key Features
Solid-State RF Power Supply: Equipped with an automated load matching technology that provides a stable 500–1600W output power. The RF generator operates with an output stability of <0.1%, supporting reliable plasma generation and sample excitation.
Echelle Grating Optical System: Employs a medium step grating (52.67 lp/mm) and an ultraviolet fused silica prism cross-dispersion method. This fixed optical design provides a high resolution of <0.0068nm at 200nm.
Automated Sample Introduction: Features a 12-rotor, 4-channel peristaltic pump where the speed can be adjusted to control the flow rate. Carrier gas, plasma gas, and auxiliary gas flows are accurately regulated by integrated mass flow controllers.
Large-Area CCD Detector: Incorporates a 27.6mm × 27.6mm CCD detector containing 1024 × 1024 addressing detection units. The detector is maintained at -45℃ using a three-stage semiconductor cooling system to reduce dark current and improve the signal-to-noise ratio.
Matrix Flexibility: The sample introduction module supports optional configurations, including imported high-salt atomizers or hydrofluoric acid (HF) resistant atomizers, to accommodate difficult sample matrices.
Constant Temperature Light Chamber: The optical chamber is maintained at a precise 35 ± 0.1℃ and utilizes a distributed nitrogen purging system (normal 2L/min, rapid 4L/min) to support measurement stability, especially in the deep ultraviolet range.
Specifications
Parameter
Specification
Model
ICP-6810-A2
Wavelength Range
165nm – 950nm
Optical System
Echelle grating (52.67 lp/mm) and prism cross-dispersion
Optical Resolution
< 0.0068 nm @ 200nm
Detector
CCD detector (1024 × 1024 pixels)
Detector Cooling
Three-stage semiconductor cooling to -45℃
RF Generator Power
500W – 1600W (Output stability <0.1%)
Observation Method
Bidirectional observation (vertical and horizontal)
Sample Injection
12-rotor, 4-channel peristaltic pump
Gas Flow Control
Mass flow controllers (Carrier, Plasma, Auxiliary)
Light Chamber Temp
Constant 35 ± 0.1 ℃
Purge Gas
Nitrogen (Distributed purging)
Detection Limit
Most elements are 1ppb ~ 10ppb
Short-term Stability
RSD < 1%
Long-term Stability
RSD < 1.5% @ 2 hours
Power Requirements
220VAC, 20A
Applications
The ICP-6810-A2 is utilized in diverse industrial and scientific fields for multi-element quantitative testing:
Geological and Ore Analysis: Determination of elements such as Ca, Mg, Na, Fe, Cu, Mn, Zn, Co, Ni, Au, and Ag in rock and soil extracts.
Petrochemical Industry: Testing more than 30 elements in crude oil and petroleum products, primarily evaluating Fe, Na, Mg, Ni, V, Ca, Pb, and Mo.
Metallurgy and Silicon Processing: Analyzing trace impurity elements (e.g., As, Bi, Pb, Sb, Sn) that affect the quality of metal and magnetic materials.
Environmental and Water Quality: Detecting heavy metals and elemental pollutants in wastewater, drinking water, and environmental monitoring programs.
Who is it for?
Commercial Testing Laboratories: Facilities that process high volumes of liquid samples daily and require a high-throughput instrument capable of reading all target elements within one minute.
Industrial Quality Control Departments: QC labs in the petrochemical, metallurgical, and semiconductor sectors that require stable testing across a broad dynamic range (from trace ppb levels to percentage concentrations).
Scientific Research Institutions: Organizations requiring reliable quantitative analysis for applications in clinical medicine, agricultural research, and criminal science.
Why choose it?
High Analytical Throughput: By utilizing a large-area CCD detector without moving optical components, the instrument allows the simultaneous exposure and readout of multiple analytical spectral lines, significantly accelerating testing speed.
Adaptable to Complex Matrices: The ability to configure the system with high-salt or HF-resistant atomizers ensures that challenging sample digestions (such as those using hydrofluoric acid for silicates) can be processed without damaging the introduction hardware.
Instrument Stability: The integration of a solid-state RF power supply, mass flow controllers for precise gas management, and a thermally stabilized optical chamber supports a short-term stability RSD of <1% and a long-term stability of <1.5% over two hours.
Certified Qualifications: The instrument complies with major international product qualifications and safety standards, holding certifications including the European CE mark, North American ETL, FCC, RoHS, China CQC, and 3C.
Packaging, Installation & Operation
Packaging & Shipping
After unpacking the instrument, compare the contents with the packing list and check that all items are present and undamaged.
Installation & Setup
Place the spectrophotometer in a suitable location away from direct sunlight, dust, corrosive gases, and strong vibrations. For stable operation, keep it away from strong magnetic or electrical fields and electrical devices that may generate high-frequency fields.
Keep the area around and underneath the instrument clear to allow adequate airflow.
Use the appropriate power cord and connect the instrument to a grounded outlet.
Initial Operation
Turn on the spectrophotometer and allow it to warm up for the recommended time before taking measurements. Follow the instrument’s operating instructions to complete any required system initialization, calibration, or wavelength verification before measurement.
Exhibition
FAQ (Frequently Asked Questions)
The instrument covers a full-spectrum wavelength range from 165nm to 950nm, supporting the analysis of elements that emit in the deep ultraviolet and visible-near-infrared regions.
Yes. The sample introduction system can be configured with an optional HF-resistant atomizer, making it suitable for analyzing geological or semiconductor samples digested in hydrofluoric acid.
No. The ICP-6810-A2 utilizes a medium step echelle grating and prism cross-dispersion method with no moving optical components. Wavelength calibration and spectral line identification are handled automatically by the instrument’s software.
The carrier gas, plasma gas, and auxiliary gas are all controlled by integrated mass flow controllers. This ensures precise, automated flow rates to maintain plasma stability during injection.
The CCD detector is cooled using an efficient three-stage semiconductor cooling module, which lowers the operating temperature to -45℃ to reduce dark current and improve detection limits.