Key Features

  • Solid-State RF Power Supply: Equipped with an automatic load matching technology that provides a 500–1600W output power. The RF generator operates with an output power stability of <0.1%, supporting stable plasma generation and efficient sample excitation.
  • Automated Sample Introduction: Features a 12-rotor, 4-channel peristaltic pump with adjustable speed control. Carrier gas, plasma gas, and auxiliary gas flows are regulated by integrated mass flow controllers to ensure stable gas delivery.
  • High-Resolution Optical System: Employs a medium step grating (52.67 lp/mm) and a Corning ultraviolet fused silica prism. The fixed optical design provides a high resolution of <0.0068nm at 200nm.
  • Large-Area CCD Detector: Incorporates a 27.6mm × 27.6mm CCD detector containing 1024 × 1024 addressing detection units. The detector is maintained at -45℃ using a three-stage semiconductor cooling system to reduce dark current.
  • Matrix Adaptability: The sample introduction module can be equipped with optional components, including imported high-salt atomizers or hydrofluoric acid (HF) resistant atomizers, to accommodate various challenging sample matrices.
  • Constant Temperature Light Chamber: The precision optical chamber is maintained at a constant 35 ± 0.1℃ and utilizes a distributed nitrogen purging system (normal 2L/min, rapid 4L/min) to support stability during ultraviolet region measurements.

Specifications

ParameterSpecification
ModelICP-6810-A1
Wavelength Range165nm – 950nm
Optical SystemEchelle grating (52.67 lp/mm) and prism cross-dispersion
Optical Resolution< 0.0068 nm @ 200nm
DetectorCCD detector (1024 × 1024 pixels)
Detector CoolingThree-stage semiconductor cooling to -45℃
RF Generator Power500W – 1600W (Output stability <0.1%)
Sample Injection12-rotor, 4-channel peristaltic pump
Gas Flow ControlMass flow controllers (Carrier, Plasma, Auxiliary)
Light Chamber TempConstant 35 ± 0.1 ℃
Purge GasNitrogen (Distributed purging)
Liquid Content Range0.01ppm ~ several thousand ppm
Solid Content Range0.001% ~ 70%
Short-term StabilityRSD < 1%
Long-term StabilityRSD < 1.5% @ 2 hours
Power Requirements220VAC, 20A

Applications

The ICP-6810-A1 is utilized in industrial, environmental, and scientific fields for multi-element quantitative testing:

  • Geological and Ore Analysis: Determination of elements such as Ca, Mg, Na, Fe, Cu, Mn, Zn, Co, Ni, Au, and Ag in rock and soil extracts.
  • Petrochemical Industry: Testing more than 30 elements in crude oil and petroleum products, including Fe, Na, Mg, Ni, V, Ca, Pb, and Mo.
  • Metallurgy and Silicon Processing: Analyzing trace impurity elements (e.g., As, Bi, Pb, Sb, Sn) that impact the quality of metal and magnetic materials.
  • Environmental and Water Quality: Detecting heavy metals and elemental pollutants in wastewater, drinking water, and environmental compliance testing.

Who is it for?

  • Commercial Testing Laboratories: Facilities that process large volumes of liquid samples daily and require an instrument capable of rapid simultaneous readout of all elements.
  • Industrial Quality Control Departments: QC labs in the petrochemical, metallurgical, and semiconductor sectors that require stable testing across a broad dynamic range.
  • Research Institutions: Organizations requiring reliable quantitative analysis for applications in agricultural research, criminal science, and clinical medicine.

Why choose it?

  • Fast Analytical Speed: By utilizing a large-area CCD detector without moving optical components, the instrument performs simultaneous exposure and readout. The readout time of a single spectral line is less than 2ms, significantly accelerating the analysis process.
  • Configurable for Complex Matrices: The ability to configure the system with high-salt or HF-resistant atomizers ensures that samples digested in hydrofluoric acid or featuring high dissolved solids can be processed effectively.
  • Instrument Stability: The integration of a solid-state RF power supply, mass flow controllers for precise gas management, and a thermally stabilized optical chamber helps maintain a short-term stability RSD of <1% and a long-term stability of <1.5% over two hours.
  • Certified Qualifications: The instrument complies with major international product qualifications and safety standards, holding certifications including the European CE mark, North American ETL, FCC, RoHS, China CQC, and 3C.

Packaging, Installation & Operation

Packaging & Shipping

After unpacking the instrument, compare the contents with the packing list and check that all items are present and undamaged.

Installation & Setup

Place the spectrophotometer in a suitable location away from direct sunlight, dust, corrosive gases, and strong vibrations. For stable operation, keep it away from strong magnetic or electrical fields and electrical devices that may generate high-frequency fields.

Keep the area around and underneath the instrument clear to allow adequate airflow.

Use the appropriate power cord and connect the instrument to a grounded outlet.

Initial Operation

Turn on the spectrophotometer and allow it to warm up for the recommended time before taking measurements. Follow the instrument’s operating instructions to complete any required system initialization, calibration, or wavelength verification before measurement.

products-Packaging-Shipping.

Exhibition

products-Exhibition.

FAQ (Frequently Asked Questions)

The instrument covers a full-spectrum wavelength range from 165nm to 950nm, supporting the analysis of elements that emit in the ultraviolet and visible-near-infrared regions.

Yes. The sample introduction system can be equipped with an optional HF-resistant atomizer, making it suitable for analyzing geological or semiconductor samples digested in hydrofluoric acid.

No. The ICP-6810-A1 utilizes a medium step echelle grating and prism cross-dispersion method with no moving optical components. Wavelength calibration and spectral line identification are handled automatically by the instrument’s software algorithms.

The carrier gas, plasma gas, and auxiliary gas are controlled by integrated mass flow controllers. This ensures precise flow rates to maintain plasma stability during testing.

The CCD detector is cooled using a three-stage semiconductor cooling module, which lowers the operating temperature to -45℃ to reduce dark current and improve signal detection.

×