Performance characteristics

RF Power technical parameter

Circuit type: solid-state RF power supply, with function of automatch
Frequency: 27.12MHz±0.05%
Frequency Stability: <0.1%
Power Output: 500—1600W
Power Output Stability: <0.3%
Escaped RF radiation: 30cm away from the instrument, electric field: E﹤2V/m

Sampling System Technical Parameter

Output working coil inner diameter :25mm
Torque tube: Three concentric, external diameter 20mm
Coaxial nebulizer:Outer diameter 6mm
Double barrel atomizing chamber: Outer diameter 34mm

Gas Flow Controls

Plasma Argon Flowmeter:(100-1000)L/h (1.6-16L/min)
Auxiliary Argon Flowmeter:(10-100)L/h (0.16-1.66L/min)
Carrier Argon Flowmeter(10-100)L/h (0.16-1.66L/min)
Pressure Maintaining Valve(0-0.4MPa)
Cooling Water:Temperature: 20-25℃, Rate of Flow>5L/min, Hydraulic Pressure>0.1MPa

Spectrometer

Optics:Czerny-Turner type
Focal length: 1000 mm
Grating: Ion Beam Etching Holographic Grating, 3600L/mm or 2400L/mm
Reciprocal linear dispersion: 0.26nm/mm
Resolution:≤ 0.007nm (3600 line grating);≤0.015nm (2400line grating)
Wavelength range: 3600 line grating:(190nm~500) nm;2400 line grating: (190nm~800) nm
Minimum pace of stepping motor: ≤0.0006 nm
Exit Slit:12μm;Entrance Slit:10μm

Photoelectric Converter Performance

Photomultiplier tube specification: R293/R928
Negative HV on PMT: 0-1000V;Stability: <0.05%

Detection Limits (μg/L)

Element WL(nm) Limit Element WL(nm) Limit
La 408.672 <3.0 Cr 267.716 <5.0
Ce 413.765 <5.0 Al 396.152 <5.0
Pr 414.311 <5.0 Zr 343.823 <5.0
Nd 401.225 <5.0 Ag 328.068 <3.0
Sm 360.946 <10.0 Sr 407.771 <1.0
Eu 381.967 <1.0 Au 242.795 <5.0
Gd 342.247 <10.0 Pt 265.945 <5.0
Tb 350.917 <3.0 Pd 340.458 <5.0
Dy 353.170 <3.0 Ir 224.268 <10.0
Ho 345.600 <3.0 Rh 343.489 <10.0
Er 337.271 <3.0 Ru 240.272 <5.0
Tm 313.126 <3.0 Ba 455.403 <1.0
Yb 369.419 <1.0 As 228.812 ≤15
Lu 261.541 <3.0 Sb 206.833 ≤15
Y 371.030 <1.0 Bi 223.061 ≤10
Sc 335.373 <1.0 Hg 253.652 ≤15
Ta 226.230 <5.0 Pb 220.353 ≤15
Nb 313.340 <5.0 Ga 294.364 ≤10
Mn 257.610 <3.0 Se 203.985 ≤10
Mg 279.553 <1.0 Sn 242.949 ≤20
B 249.773 <10.0 Te 214.281 ≤10
Zn 213.856 <3.0 Ta 226.230 ≤5.0
Co 228.616 <3.0 Th 283.730 ≤10
Si 251.611 <10.0 Tl 276.787 ≤30
Ni 232.003 <5.0 Re 227.525 ≤5
Cd 226.502 <3.0 Ge 209.426 ≤15
Fe 239.562 <3.0 Os 225.585 ≤1
Ca 393.366 <1.0 W 207.911 ≤10
Mo 281.615 <5.0 Cu 324.754 <3.0
V 310.230 <5.0 Li 670.784 ≤3
Be 313.041 <1.0 Na 588.995 ≤20
Ti 334.941 <3.0 K 766.490 ≤60

Certifications

Product Qualification: European safety qualification mark CE, North American safety certification ETL, Federal Communications Commission FCC, European electrical safety certification RoHS, China Quality Certification Center CQC, China national mandatory product certification 3C

Packaging & Shipping

After carefully unpacking the contents, check the materials with the packing list (page 4) to ensure that you have received everything in good condition.

Place the instrument in a suitable location away from direct sunlight.  In order to have the best performance from your instrument, keep it as far as possible from any strong magnetic or electrical fields or any electrical device that may generate high-frequency fields. Set the unit up in an area that is free of dust, corrosive gases and strong vibrations.

Remove any obstructions or materials that could hinder the flow of air under and around the instrument.

Use the appropriate power cord and plug into a grounded outlet.

Turn on your spectrophotometer. Allow it to warm up for 15 minutes before taking any readings. We suggest you then do the Calibrate System with the Search 656.1nm to set the wavelength to the deuterium lamp emission line.